User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2618 DATE: 13-May-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 90% NUMBER OF TESTS: 29 NUMBER OF LINES: 176 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 6060B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ D N 1.002 ASK- R P J F W 1.003 HEAD {DC-VOLTAGE ACCURACY TEST} 1.004 DISP Four "C" size batteries must be installed. 1.004 DISP Verify the UUT backlit by putting the UUT in a dark 1.004 DISP area ( applicable on PM 2618/3XX ). 1.004 DISP Set the function switch to V-DC . 1.004 DISP Press key until UUT displays "000.0 mV *". 1.004 DISP Connect a banana cable to UUT OHM-Hz-V-mA (red terminal) 1.004 DISP and O (black terminal). Place a short across the end of 1.004 DISP the cable. Then press key on UUT. 1.005 DISP Remove the short and connect the cable to MULTIFUNCTION 1.005 DISP CALIBRATOR Output. 1.006 5100 0.1V S 2W 1.007 MEMI Please enter UUT reading: Example "###.#" 1.008 MEM/ 1000 1.009 MEMC .1000V 0.0003U 2.001 DISP Set UUT to 10V range by pressing key once. 2.002 5100 10V S 2W 2.003 MEMI Please enter the UUT reading: Example "##.###" 2.004 MEMC 10.000V 0.009U 3.001 5100 -10V S 2W 3.002 MEMI Please enter the UUT reading: Example "##.###" 3.003 MEMC -10.000V 0.009U 4.001 DISP Set UUT to 100V range by pressing key once. 4.002 5100 100V S 2W 4.003 MEMI Please enter the UUT reading: Example "###.##" 4.004 MEMC 100.00V 0.09U 5.001 DISP Set UUT to 1000V range by pressing key once. 5.001 DISP CAUTION : [27][91]7m HIGH VOLTAGE ! [27][91]m 5.002 5100 900V S 2W 5.003 MEMI Please enter the UUT reading: Example "####.#" 5.004 MEMC 900.0V 1.1U 6.001 5100 0V S 2W 6.002 5100 * S 6.003 RSLT = 6.004 HEAD {AC-VOLTAGE ACCURACY TEST} 6.005 DISP Set the function switch to V-AC . 6.005 DISP Set UUT to 1V range by pressing key until UUT 6.005 DISP displays "~ 000.0 mV *". 6.006 5100 0.1V 50H S 2W 6.007 MEMI Please enter UUT reading: Example "###.#" 6.008 MEM/ 1000 6.009 MEMC .1000V 0.0014U 50H 7.001 5100 0.1V 50kH S 2W 7.002 MEMI Please enter UUT reading: Example "###.#" 7.003 MEM/ 1000 7.004 MEMC .1000V 0.02U 50kH 8.001 DISP Set UUT to 10V range by pressing key once. 8.002 5100 10V 50H S 2W 8.003 MEMI Please enter the UUT reading: Example "##.###" 8.004 MEMC 10.000V 0.05U 50H 9.001 5100 10V 50kH S 2W 9.002 MEMI Please enter the UUT reading: Example "##.###" 9.003 MEMC 10.000V 0.65U 50kH 10.001 DISP Set UUT to 100V range by pressing key once. 10.002 5100 100V 50H S 2W 10.003 MEMI Please enter the UUT reading: Example "###.##" 10.004 MEMC 100.00V 0.5U 50H 11.001 5100 100V 1kH S 2W 11.002 MEMI Please enter the UUT reading: Example "###.##" 11.003 MEMC 100.00V 0.5U 1kH 12.001 DISP Set UUT to 1000V range by pressing key once. 12.001 DISP CAUTION : [27][91]7m HIGH VOLTAGE ! [27][91]m 12.002 5100 900V 50H S 2W 12.003 MEMI Please enter the UUT reading: Example "####.#" 12.004 MEMC 900.0V 3.9U 50H 13.001 5100 900V 1kH S 2W 13.002 MEMI Please enter the UUT reading: Example "####.#" 13.003 MEMC 900.0V 3.9U 1kH 14.001 5100 * S 14.002 RSLT = 14.003 HEAD {RESISTANCE ACCURACY TEST} 14.004 DISP Set the function switch to . 14.004 DISP Press the key until UUT displays " OL. *". 14.004 DISP Disconect the cable from MULTIFUNCTION CALIBRATOR 14.004 DISP Output and place a short across the cable. 14.004 DISP Press key then remove the short from the cable. 14.004 DISP Connect the cable to MULTIFUNCTION CALIBRATOR Output. 14.005 RESF 1000Z S 2W 14.006 MEMI Please enter UUT reading: Example "####.#" 14.007 MEMC 1000.0Z 3U 15.001 DISP Set UUT to 10 KOhm range by pressing key once. 15.002 RESF 10kZ S 2W 15.003 MEMI Please enter UUT reading: Example "##.###" 15.004 MEMC 10.000KZ 0.03U 16.001 DISP Set UUT to 100 KOhm range by pressing key once. 16.002 RESF 100kZ S 2W 16.003 MEMI Please enter UUT reading: Example "###.##" 16.004 MEMC 100.00KZ 0.3U 17.001 DISP Set UUt to 1 MOhm range by pressing key once. 17.002 RESF 1MZ S 2W 17.003 MEMI Please enter UUT reading: Example "####.#" 17.004 MEM/ 1000 17.005 MEMC 1.0000MZ 0.003U 18.001 DISP Set UUT to 10 MOhm range by pressing key once. 18.002 RESF 10MZ S 2W 18.003 MEMI Please enter UUT reading: Example "##.###" 18.004 MEMC 10.000MZ 0.05U 19.001 RESF * S 19.002 RSLT = 19.003 HEAD {CONTINUITY AND BEEPER TEST} 19.004 DISP Set the function switch to . 19.005 RESF 1Z S 2W 19.006 EVAL Does the UUT displays "00X.X" and activates the BEEPER ? 20.001 RESF * S 20.002 RSLT = 20.003 HEAD {FREQUENCY ACCURACY TEST} 20.004 DISP Set the function switch to . 20.004 DISP Disconnect the cable from MULTIFUNCTION CALIBRATOR 20.004 DISP Output and connect it to the 6060B RF Output. 20.005 6060 19kH 1VP -D S 20.006 MEMI Please enter UUT reading: Example "##.###" 20.007 MEMC 19.000KH 0.019U 21.001 6060 199kH 1VP -D S 21.002 MEMI Please enter UUT reading: Example "###.##" 21.003 MEMC 199.00KH 0.199U 22.001 6060 * S 22.002 RSLT = 22.003 HEAD {DC-CURRENT ACCURACY TEST} 22.004 DISP Set the function switch to . Press key until 22.004 DISP UUT displays "00.00 mA *". 22.004 DISP Move the cable to MULTIFUNCTION CALIBRATOR Output from 22.004 DISP the 6060 RF Output. 22.005 5100 20mA S 2W 22.006 MEMI Please enter UUT reading: Example "##.##" 22.007 MEMC 20.00mA 0.1U 23.001 5100 -20mA S 2W 23.002 MEMI Please enter UUT reading: Example "-##.##" 23.003 MEMC -20.00mA 0.1U 24.001 DISP Set UUT to 200mA range by pressing key once. 24.002 5100 200mA S 2W 24.003 MEMI Please enter UUT reading: Example "###.#" 24.004 MEMC 200.0mA 1U 25.001 DISP Move the red terminal to <10A> input from V-Ohm input. 25.001 DISP Press key until UUT displays "0.000 A *". 25.001 DISP CAUTION : [27][91]7m HIGH CURRENT ! [27][91]m 25.002 5100 1A S 2W 25.003 MEMI Please enter UUT reading: Example "#.###" 25.004 MEMC 1.000A 0.006U 26.001 5100 * S 26.002 RSLT = 26.003 HEAD {AC - CURRENT ACCURACY TEST} 26.004 DISP Set function switch to . 26.004 DISP Press key until UUT displays "~ 0.000 A *". 26.004 DISP CAUTION : [27][91]7m HIGH CURRENT ! [27][91]m 26.005 5100 1A 50H S 2W 26.006 MEMI Please enter UUT reading: Example "#.###" 26.007 MEMC 1.000A 0.012U 50H 27.001 5100 * S 27.002 DISP Move the red terminal to input from <10A> 27.002 DISP input. 27.002 DISP Press key until the UUT displays "~00.00 mA *". 27.003 5100 20mA 50H S 2W 27.004 MEMI Please enter UUT reading: Example "##.##" 27.005 MEMC 20.00mA 0.18U 50H 28.001 DISP Set UUT to 200mA range by pressing once. 28.002 5100 200mA 50H S 2W 28.003 MEMI Please enter UUT reading: Example "##.##" 28.004 MEMC 200.0mA 1.8U 50H 29.001 5100 * S 29.002 RSLT = 29.003 HEAD THE END